8-900 Flow-thru Conductivity Electrode with 1/8" Fittings
8-900 Flow-thru Conductivity Electrode with 1/8" Fittings
SPECIFICATIONS
TOTAL LENGTH
3.2 cm LEAD LENGTH
1 m
BODY (Outer Diameter)
10 mm PolyCarbonate
Barb
1/8" Polycarbonate
RESPONSE TIME:
5 to 15 seconds
DEPTH of IMMERSION:
1/8" flow-thru
RANGE:
10 uS - 100 mS
TEMPERATURE RANGE:
-5°C to 75° C
CALIBRATION
Pumping a known calibration standard (0.01 KCl) will help in determining the K for the electrode
CLEANING
Using any electrode in solutions containing protein requires the electrode be rinsed with an enzyme cleaning solution.
After each use, we recommend cleaning our electrodes with Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning solution by submerging the electrodes for a couple of minutes in order to remove all protein from the glass and reference junction.
This will prolong the useful life of the electrodes.
STORAGE
Always clean the microelectrode before storing:
- Long-term (over 2 weeks): rinse flow path with DI water and dry
- Short-term: Rince flow path with DI water