8-702 Flow-thru Reference Electrode with 1/8" Fittings

8-702 Flow-thru Reference Electrode with 1/8" Fittings

$199.00Price

8-702 Flow-thru Reference Electrode with 1/8" Fittings

  • SPECIFICATIONS

    JUNCTION                             

    Ceramic frit

    TOTAL LENGTH

    6.5 cm 

    LEAD LENGTH

    2 m

    DIAMETER of REFERENCE BARREL

    3 mm 

    MATERIAL of BARREL

    Acrylic 

    MATERIAL OF FLOW CELL:

    Acrylic

    Polysulfone

    CELL VOLUME

    0.093 ml & 1.011 ml  

    "T" FITTING

    1/8" 

    INTERNAL ELEMENT

    ELECTROLYTE

    Ag-AgCl

    3M KCI 

  • CALIBRATION

    Carefully unwind the tape and remove the probe from the protective glass tube.  The white translucent sleeve, which covers the hole, should be moved down the glass barrel slightly to ventilate the reference chamber and replaced after each use. 

     

    The use of a separate Reference Electrode is not required.

     

    The probe can be held by hand when measuring pH.  It has been standardized in two (2) pH buffers.  Follow the procedure recommended by the manufacturer of your pH meter for calibrating your pH meter with our microelectrode.

     

    OPTIMUM RESPONSE TIME will be obtained after the probe has been exercised in two (2) buffer solutions. 

     

    • Fill one beaker with a pH 4 buffer or equivalent and a second beaker with a pH 7 buffer or equivalent. 
    • Hold the microelectrode and touch the pH 4 buffer surface allowing 5 seconds for equilibration, and then touch the pH 7 buffer surface in the same manner.  Do this several times. 
    • It is not necessary to rinse the probe because the carry-over is negligible.
  • CLEANING

    Using any electrode in solutions containing protein requires the electrode be rinsed with an enzyme cleaning solution.

     

    After each use, we recommend cleaning our electrodes with Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning solution by submerging the electrodes for a couple of minutes in order to remove all protein from the glass and reference junction.

     

    This will prolong the useful life of the electrodes.

  • STORAGE

    Always clean the microelectrode before storing:

     

    • Long-term (over 4 weeks): Drain the solution and rince with DI water

     

    • Short-term: The probe can be left in any solution to keep the tip from drying out.
  • Instructions pdf