top of page
16-720 Flow-thru Carbon Dioxide Electrode with 1/16" Fittings

16-720 Flow-thru Carbon Dioxide Electrode with 1/16" Fittings

$585.00Price

16-720 Flow-thru Carbon Dioxide Electrode with 1/16" Fittings

These miniature electrodes are ideal for monitoring Carbon Dioxide flow such as water, blood, urine, physiologic fluids, etc. They feature fast response and require less than a drop.  Each microelectrode is supplied with 6 assembled membranes and a bottle of electrolyte solution.

  • SPECIFICATIONS

    TOTAL LENGTH                             

    8.6 cm

    LEAD LENGTH

    2 m  

    BODY (Outer Diameter)

    6 mm

    TIP (Outer Diameter)

    pH: 0.75 mm

    Ref + pH: 1.3 mm 

    RESPONSE TIME:

    less than 1 minute  

    DEPTH of IMMERSION:

    0.5 mm to 1 mm  

    SENSITIVITY:

    0.4 mv/mm CO2  

    TEMPERATURE RANGE:

    -5°C to 100° C  

    REFERENCE ELECTRODES

    TYPE: Solution & Electrode

     

    3M KCI & Ag-AgCI

  • CALIBRATION

    Calibration is done with two standard gases bubbled through a liquid. Typically 5% and 10% CO2 balanced Nitrogen. The probe responds slugishly to low CO2 levels, so to help with repsonse time, the amount of exposure to ambient should be limited in use.

  • CLEANING

    Using any electrode in solutions containing protein requires the electrode be rinsed with an enzyme cleaning solution.

     

    After each use, we recommend cleaning our electrodes with Terg-a-zyme (Alconox, Inc.) solution by submerging the electrodes for a couple of minutes in order to remove all protein from the glass and reference junction.

     

    This will prolong the useful life of the electrodes.

  • STORAGE

    Always clean the microelectrode before storing:

     

    • Long-term (over 2 weeks): Rince membrane and keep on electrode for protection
bottom of page